A comprehensive reference for users who want to understand and work with the spectral postprocessing capabilities of photon-counting CT in syngo.via.
syngo.CT Dual Energy – Instructions for Use
This document provides detailed guidance on spectral CT postprocessing with syngo.CT Dual Energy, including functionality relevant to photon-counting CT (PCCT). It covers QuantumPlus and QuantumPeak data and describes how spectral information can be used for image reconstruction, visualization, quantitative evaluation, and clinical applications.
PCCT-relevant topics include:
- Spectral Postprocessing (SPP) and the generation of spectral datasets for further processing.
- Monoenergetic and Monoenergetic Plus imaging for energy-dependent visualization, including improved iodine contrast and metal artifact reduction.
- Virtual non-contrast (VNC) and iodine imaging based on spectral material decomposition.
- Clinical applications including Liver VNC, Lung Analysis, Kidney Stones, Gout, Brain Hemorrhage and Bone Marrow, depending on the available spectral acquisition mode.
- Detailed information on image requirements, workflows, measurements, limitations, safety considerations and configuration of spectral postprocessing.
The document also explains how photon-counting detector data are processed within syngo.via and how spectral results can be incorporated into clinical reading and reporting workflows.
What I changed: I removed the legacy Dual Energy detail from the foreground and made the description website-oriented, with PCCT, SPP, spectral applications, and clinical workflow as the main entry points.
Prerequisite
syngo.CT Dual Energy also supports the generation of SPR (Stopping Power
Ratio) and SPP (Spectral Post Processing) images. The SPR application class
allows Rapid Results Technology to calculate SPR images in any workflow.
Quantum spectral imaging
To load Spectral Postprocessing (SPP) data from Quantum technology systems
into the CT Dual Energy workflow, the following image requirements need to
be met:
- Spectral Postprocessing (SPP) images or Vascular Spectral Postprocessing
(VSPP) images have been acquired using a scan mode with 120 kV or 140 kV
on a Quantum technology system.
- The minimum number of axial CT images is 6.
- The images have a matrix of 512 x 512.
- Depending on the application class, the series need to be acquired with or
without an iodine contrast agent.
- It is recommended that the SPP series contains a maximum of 4095 images.
Details
The Virtual Unenhanced application class allows you to visualize the contrast
agent concentration in soft tissue and non-enhanced soft tissue without the
need for an additional non-contrast scan.
The Monoenergetic Plus application class simulates images that are
equivalent to images scanned with an X-ray beam of a single energy level
(keV). By changing the energy (keV), physicians can enhance the contrast
between different materials or reduce metal artifacts.
The Bone Marrow application class allows you to visualize the bone marrow
composition based on non-enhanced CT data. The application class generates
virtual non-calcium (VNCa) images by subtracting calcium from the data sets.
The Bone Removal application classes allow you to remove bone or orthopedic
implants, such as nails, from CT angiography (CTA) data sets. The prerequisite
for bone removal is the decomposition of the material into its component parts:
blood, contrast agent, and bone.
The Brain Hemorrhage application class allows you to distinguish contrast
agent from a hemorrhage in the brain. This application class is used if the
head was scanned with a dedicated brain hemorrhage scan protocol and a
suspicious lesion is spotted in the Conventional CT images.
The Gout application class allows you to distinguish between urate, bone, bone
marrow, and contrast agent. Urate and contrast agent are highlighted with
different colors.
The Hard Plaques application class allows you to highlight calcified plaques
within large vessels even if they have CT values that are comparable to
the neighboring contrast agent. The basis for this approach is a material
decomposition into iodine contrast agent, blood, and bone.
The Heart PBV (Perfused Blood Volume) application class allows you to
visualize the contrast agent uptake in the myocardium. The isolation of the
heart is part of the preprocessing.
The Kidney Stones application class visualizes the chemical differences
between kidney stones by identifying the base material of the kidney stone, for
example, uric acid or oxalate.
The Liver VNC application class allows you to visualize the contrast agent
concentration in the liver without an additional non-contrast scan, even if
there are irregular fatty infiltrations or necrotic areas. The application class
generates virtual non-contrast (VNC) images by subtracting iodine from the
data sets.
Lung Analysis combines the following application subclasses: Lung Vessels,
which visualizes the iodine enhancement in pulmonary vessels, and Lung PBV
(Perfused Blood Volume) or Lung Mono, which visualizes and quantifies the
iodine uptake in the lung parenchyma.
A unique feature of the Dual Energy CT system is the ability for chemical
characterization of materials. The Rho/Z application class provides electron
density and effective atomic number images and thus performs a basic
characterization of the material of interest.
The following tables provide an overview of the available application classes
and functions for each scan mode: